Pipette calibration (ISO 8655)

Turn weighed water masses into accuracy and precision against the nominal volume.

V = mass × Z
Mass of each dispensed water aliquot.
ISO 8655V = mass × Z; accuracy = mean − nominal; precision = SD/CV
Calculated live

Mean 100.2 µL, accuracy +0.2%, CV 0.166%

1

Accuracy is the systematic error (mean minus nominal); precision is the CV. Compare both to your pipette class limits.

Z = 1.003 µL/mg at 20 °C. ISO 8655 calls for 10 weighings; you have 10.

Mean volume
100.2 µL
Accuracy
+0.2 µL
CV
0.166%
Z
1.003 µL/mg
Pipette calibration (ISO 8655), Aliquot